model-calys-50-bo-hieu-chuan-da-chuc-nang-aoip-viet-nam.png

CALYS 50 field multifunction calibrator is the perfect tool for advanced process maintenance and use on test bench in all industries. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process and frequency in one single instrument.

 

CALYS 50 is a basic multifunction calibrator within CALYS range. It is the perfect tool for advanced process maintenance and use on test bench in all industries. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process and frequency in one single instrument.

It provides extended functionalities (temperature simulation, scaling, steps, synthesizer, statistical functions…), easily accessible through quick menus by function.

The instrument simultaneously measures and simulates:

  • Temperature: Up to 0,014% RDG
  • Resistance: Up to 0,012% RDG, 4 KΩ range
  • Current: Up to 0,0175% RDG, 50 mA range + 24 V loop supply
  • Voltage: Up to 0,013% RDG, 50 V IN / 20 V OUT range
  • Frequency: Up to 0,005% RDG, 20 KHz range (10 KHz in simulation)

IP 54, fully protected by an antichoc rubber holster, CALYS 50 integrates "easyconnect" terminals and a wide backlite display that makes it easy to use in any severe or dark conditions.

Easy connection system

CALYS-EasyconnectConnect your probes by simply pushing on the terminal top and insert wires of up to 3 mm or 10 AWG diameter and compensated thermocouple connectors.

Wires are held tight between two brass plates ensuring thermal stability and a very good cold junction compensation for thermocouples.

This system also enables 4 mm banana plugs and security connectors to be connected on the terminal top.

 

CALYS series, 4 models from basic use to advanced performances

Specifications CALYS 50 CALYS 75 CALYS 100 CALYS 150
Top accuracy 200 ppm 130 ppm 50 ppm
Temperature accuracy Thermocouples (14)
RTDs (12)
0.013% RDG for Tc K
0.012% RDG
0.01% RDG for Tc K
0.01% RDG
0.005% RDG for Tc K
0.006% RDG
DC current
+ Loop supply 24 V
Range
Accuracy
50 mA
0.0175% RDG
100 mA
0.007% RDG
DC voltage Range
Accuracy
50 V IN / 20 V OUT
0.013% RDG
50 V
0.010% RDG
50 V
0.005% RDG
Frequency Range
Accuracy
20 KHz IN / 10 KHz OUT
0.005% RDG
100 KHz
0.01% RDG
Resistance Range
Accuracy
4000 Ω
0.012% RDG
4000 Ω
0.010% RDG
50 KΩ
0.006% RDG
Pressure Range
Accuracy
  Relative pressure: 30 bar / Absolute pressure: 1,000 bar
0.05% RDG
Compliance to standards       21 CFR Part 11
    NADCAP Heat treatment AMS 2750
Additional functions Advanced data exploitation: Scaling, relative measurement, simulation of ramps and steps, synthetizer, square root, statistical functions
Transmitter function
Additional functions   Switch test
Calibration of transmitters
Additional functions     Comparison calibration
HART: Digital calibration and data transfer
Calibration of thermistors
Software   DATACAL calibration software for configuration and data management
Memory   10,000 data stored and recalled on screen as curve or list

 

https://www.aoip.com/product/field-multifunction-instruments/calys-50-3/