Model : CALYS 100 Bộ hiệu chuẩn Atex AOIP VIET NAM
Hãng sản xuất: AOIP Việt Nam
CALYS 100 is a field precision documenting multifunction calibrator within CALYS range. It is the perfect tool for advanced process maintenance and use on test bench in all industries. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process, pressure and frequency in one single instrument.
Providing extended functionalities (temperature simulation, scaling, steps, synthesizer, statistical functions…), CALYS 100 makes advanced data exploitation and full data traceability easier, as well as quick access to functions by menus.
The instrument simultaneously measures and simulates:
- Temperature: Up to 0,012% RDG
- Resistance: Up to 0,010% RDG, 4 KΩ range
- Current: Up to 0,012% RDG, 50 mA range + 24 V loop supply
- Voltage: Up to 0,010% RDG, 50 V range
- Frequency: Up to 0,005% RDG, 20 KHz range (10 KHz in simulation)
- Pressure: When used with an external pressure module (ref. ACL433), CALYS 100 can measure and simulate pressure (comparison calibration with a pressure pump).
Using this user-friendly instrument, calibration tasks can be quickly carried out over the whole process chain. Take the 900 g documenting process calibrator to the field with you during the whole week with 10 calibration procedures stored in the device. Run the procedure after connecting the probes to the instrument (Easy connect system®) and save the results for onsite easy and quick calibration.
Back to the office, you can then upload the data on a computer in order to issue customized calibration certificates with dedicated calibration software DATACAL.
IP 54, fully protected by an antichoc rubber holster, CALYS 100 integrates "easyconnect" terminals and a wide backlite display that makes it easy to use in any severe or dark conditions.
Easy connection system®
Connect your probes by simply pushing on the terminal top and insert wires of up to 3 mm or 10 AWG diameter and compensated thermocouple connectors.
Wires are held tight between two brass plates ensuring thermal stability and a very good cold junction compensation for thermocouples. This system also enables 4 mm banana plugs and security connectors to be connected on the terminal top.
CALYS series, 4 models from basic use to advanced performances
Specifications | CALYS 50 | CALYS 75 | CALYS 100 | CALYS 150 | |
Top accuracy | 200 ppm | 130 ppm | 50 ppm | ||
Temperature accuracy | Thermocouples (14) RTDs (12) | 0.013% RDG for Tc K 0.012% RDG | 0.01% RDG for Tc K 0.01% RDG | 0.005% RDG for Tc K 0.006% RDG | |
DC current + Loop supply 24 V | Range Accuracy | 50 mA 0.0175% RDG | 100 mA 0.007% RDG | ||
DC voltage | Range Accuracy | 50 V IN / 20 V OUT 0.013% RDG | 50 V 0.013% RDG | 50 V 0.010% RDG | 50 V 0.005% RDG |
Frequency | Range Accuracy | 20 KHz IN / 10 KHz OUT 0.005% RDG | 100 KHz 0.01% RDG | ||
Resistance | Range Accuracy | 4000 Ω 0.012% RDG | 4000 Ω 0.010% RDG | 50 KΩ 0.006% RDG | |
Pressure | Range Accuracy | Relative pressure: 30 bar / Absolute pressure: 1,000 bar 0.05% RDG | |||
Compliance to standards | 21 CFR Part 11 | ||||
NADCAP Heat treatment AMS 2750 | |||||
Additional functions | Advanced data exploitation: Scaling, relative measurement, simulation of ramps and steps, synthetizer, square root, statistical functions Transmitter function | ||||
Additional functions | Switch test Calibration of transmitters | ||||
Additional functions | Comparison calibration HART: Digital calibration and data transfer Calibration of thermistors | ||||
Software | DATACAL calibration software for configuration and data management | ||||
Memory | 10,000 data stored and recalled on screen as curve or list |
https://www.aoip.com/product/field-multifunction-instruments/calys-100/